The Eppendorf Liquid Handling Instrument Portfolio As the inventor of the microliter system, we at Eppendorf have over fifty years of experience in precise manual and automatic pipetting/dispensing and transferring of the smallest quantities of liquids. Download document: Certificate. Containment Testing of Rotor FA-45-12-17 in the Eppendorf 5427/R Bench Top Centrifuge. Operating manual – Centrifuge 5427 R. Find EPPENDORF 5403 - Forums, Questions, Troubleshooting, Discussions at LabWrench - Productivity, Performance, and Flexibility - Eppendorf 5403 Centrifuge. Centrifuga Eppendorf Manual We introduced our very first centrifuge in 1964 as an integral part of the Eppendorf Microliter System, and revolutionized life science. Eppendorf Centrifuge 5804/5804 R with its renowned quality and reliability offers you the most cost efficient solution for your medium throughput applications–now and in the future. Download document: Certificate - Material Safety Data Sheet (MSDS) – Pivot. Operating manual – Centrifuge 58XX family. Document size: 7.1 MB.
Troubleshooting
Centrifuge 5418 / 5418 R
English (EN)
Centrifuge 5418 / 5418 R
English (EN)
32
7.2
Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. If necessary, repeat centrifugation.
2. If necessary, repeat centrifugation.
Symptom/message Cause
Remedy
LID ERROR (5418)/
LID (5418 R)
LID (5418 R)
Centrifuge lid cannot be locked.
Try again to close centrifuge lid.
LID ERROR (5418)/
LID (5418 R)
LID (5418 R)
Centrifuge lid cannot be released.
1. Switch the centrifuge off and back on.
2. Press the
2. Press the
open key.
If the error occurs again:
1. Switch off centrifuge.
2. Activate the emergency lid release
2. Activate the emergency lid release
Emergency lid release on p. 33).
LID ERROR (5418)/
LID (5418 R)
LID (5418 R)
Centrifuge lid must not be released
during a run.
during a run.
Wait for rotor to stop.
INT
Mains/power failure during a run.
Check the mains connection.
NO RPM (5418)/
Error
Error
3 (5418 R)
Error in the speed measuring system or
drive overheated.
drive overheated.
Leave the device switched on until the
error message disappears (10 s or 6
min).
error message disappears (10 s or 6
min).
Error 5
(only for 5418 R)
(only for 5418 R)
Prohibited opening of lid or lid switch is
defective during a run.
defective during a run.
1. Wait for rotor to stop.
2. Open and close again the lid of the
2. Open and close again the lid of the
device.
3. Repeat the run.
Err 6 (5418)/
Error
Error
6 (5418 R)
Drive fault.
Repeat the run.
If this error message appears again,
switch centrifuge off and back on
again after > 20 s.
switch centrifuge off and back on
again after > 20 s.
Error 6
(only for 5418 R)
(only for 5418 R)
Drive overheated.
Allow the drive to cool down for at
least 15 min.
least 15 min.
Eppendorf Centrifuge 5403 Manual Download Free Software Software
Err 7 (5418)/
Error
Error
7 (5418 R)
Major deviation in the speed control.
1. Wait for rotor to stop.
2. Tighten rotor.
2. Tighten rotor.
Err 8 (5418)/
Error
Error
8 (5418 R)
Drive fault.
1. Wait for rotor to stop.
2. Repeat the run.
2. Repeat the run.
Eppendorf Centrifuge 5403 Manual
Err 9 to 17 (5418)/
Error
Error
9 to
17 (5418
R)
Electronics error.
![Centrifuge Centrifuge](/uploads/1/2/6/4/126442172/283464593.jpg)
Switch centrifuge off and back on
again after > 20 s.
again after > 20 s.
Error 18
(only for 5418 R)
(only for 5418 R)
Too high temperature deviation from set
value in the rotor chamber.
value in the rotor chamber.
Check the settings.
Check unhindered air circulation
through the air slots.
through the air slots.
Eppendorf 5430r Centrifuge Manual
Thaw ice or switch off device and
allow it to cool down.
allow it to cool down.